T24-SA Embedded Strain Gauge Digitiser
|
| (Specification with 1000R bridge, 2.5mV/V, at 3V supply at 25 Deg C) |
Measurement
|
| |
Min |
Typical |
Max |
Units |
| Strain Gauge Excitation Voltage |
4. 5 |
5 |
5.25 |
VDC |
| Strain Gauge Drive Capability |
85 |
- |
5000 |
Ohms |
| Strain Gauge Sensitivity (FR) |
3.2 |
3.3 |
3.4 |
+/-mV/V |
| Offset Temperature Stability |
|
1 |
4 |
ppm/C |
| Gain Temperature Stability |
|
3 |
5 |
ppm/C |
| Offset Stability with Time |
|
20 |
80 |
ppm of FR (1) |
| Gain Stability with Time |
|
|
30 |
ppm of FR (2) |
| Non Linearity before Linearization |
|
5 |
25 |
ppm of FR |
| Noise Free where Sample Time < 10mS |
|
50,000 / 15.5 |
|
Resolution/Bits |
| Noise Free where Sample Time < 50mS |
|
65,000 / 16 |
|
Resolution/Bits |
| Noise Free where Sample Time < 100mS |
|
150,000 / 17.25 |
|
Resolution/Bits |
| Noise Free where Sample Time < 1000mS |
|
250,000 / 18 |
|
Resolution/Bits |
| Noise Free where Sample Time > 1000mS |
|
400,000 / 18.75 |
|
Resolution/Bits |
Electrical
|
| |
Min |
Typical |
Max |
Units |
| Power Supply voltage |
2.1 |
3.0 |
3.6 |
V dc |
| Power Supply ripple |
|
|
50 |
mV ac pk-pk |
Power Supply current
|
| |
Min |
Typical |
Max |
Units |
| Normal Mode (1K Bridge) |
|
60 |
65 |
mA |
| Standby / Low power mode |
|
5 |
20 |
uA |
Environmental
|
| |
Min |
Typical |
Max |
Units |
| Operating temperature range |
-40 |
|
+85 |
C |
| Storage temperature |
-40 |
|
+85 |
C |
| Humidity |
0 |
|
95 |
%RH |
Telemetry Acquisition Module and Handheld
|
T24-HS Handheld Reader
|
Electrical
|
| |
Min |
Typical |
Max |
Units |
| Power Supply voltage |
2.4 |
3.0 |
3.6 |
V dc |
Power Supply
|
| |
Min |
Typical |
Max |
Units |
| Active |
|
35 |
40 |
mA |
| Low power mode |
|
120 |
160 |
uA |
| Estimated Battery life (using 2Ahr batteries) |
| Standby mode (Powered off) |
|
1.5 |
|
Years |
| Continuous operation |
|
40 |
|
Hours |
| Use a pair of alkaline AA batteries. Due to the higher voltage requirements of this device NiMh and NiCad batteries are not recommended. |
Environmental
|
| |
Min |
Typical |
Max |
Units |
| IP rating |
|
IP65 |
|
|
| Operating temperature range |
-10 |
|
+50 |
C |
| Storage temperature |
-40 |
|
+85 |
C |
| Humidity |
0 |
|
95 |
%RH |
Module Transmits & Receives giving
|
- Full error detection and correction
- Ability to switch to low power modes
- Calibration & configuration via radio telemetry
- Ability to be switched from sleep to operating mode via radio
- Calibration stored within the module