| Range (FSO) |
±10” W.G (±25 mbar or ±0.36 psi) |
| Maximum Overload Range |
110% FSO |
| Maximum Port Line Pressure |
100psi (with a differential pressure not exceeding 0.36psi) |
| Output Sensitivity |
4mV/V nominally |
| Zero Offset |
< ±2% FSO |
| Accuracy |
< ±0.5% FSO |
| Repeatability |
< ±0.2% FSO |
| Excitation Voltage |
10vdc maximum (5Vdc recommended) |
| Bridge Resistance |
700Ω |
| Insulation Resistance |
> 100MΩ at 50vdc |
| Operating Temperature Range |
-30°C to +100°C |
| Compensated Temperature Range |
+10°C to +60°C |
| Zero Temperature Coefficient |
< ±0.05% FSO/°C |
| Span Temperature Coefficient |
< ±0.05% FSO/°C |
| Environmental Protection Level |
IP66 |
| Connection Type |
Integral MIL-C-26482 Bayonet Lock Connector |
| Wiring Connections |
| Pin A |
Positive Signal |
| Pin B |
Positive Excitation |
| Pin C |
Negative Signal |
| Pin D |
Negative Excitation |
|